Indstillinger for USC 5000-strømudgang
Indstilling 1 for USC 5000-strømudgang
Figur 23 Indstillinger for USC 5000-strømudgang
* Enheder, der er valgt i sektionen "Default settings of the transmitter" (Standardindstillinger for transmitteren")
Output
USC
I out 1
USC
I out 2
USC
Function
Off
I OUT 1
On
Off
081R9412.02.01
Value
pH
I OUT 1
pH
Temperature
Value at 4 mA
+000.00 mg/l
I OUT 1
Range: 0-999
Value at 20 mA
+999.00 mg/l
I OUT 1
Range: 0-999
Output when not
measuring 20 mA
I OUT 1
Time const.
040. s
I OUT 1
030-300s
Value at 4 mA
+000.00 °C
I OUT 1
Value at 20 mA
+999.00 °C
I OUT 1
Range: -10 - 999
Range: -10 - 999
20 mA
4 mA
Hold
Normal
Output during
error Normal
I OUT 1
23 mA
3.8 mA
Normal
*
*
**
*
*