Hach-Lange BUHLER 2000 Water sampler User Manual
Page 17
Advertising
Operation
Page 17
Manual BÜHLER 2000/1029
Graphical representation of sampling modes TIME-FLOW-EVENT
Sampling modes
Flow graph
Time-
proportional
Flow-
dependent
Event signal
Event-
proportional
A sample is extracted in fixed
time intervals (e.g. every 10
minutes) with a fixed sample
volume (e.g. 50 ml).
A sample is extracted in
variable time intervals
(depending on the flow (Q))
with a fixed sample volume.
The sampler is waiting for an
event (e.g. the exceeding of
a pH limit value). A sample is
extracted in fixed time
intervals (e.g. every 10
minutes) with a fixed volume
(e.g. 50 ml) as long as an
event is present.
Advertising