Page 17

Manual BÜHLER 2000/1029

Graphical representation of sampling modes TIME-FLOW-EVENT

Sampling modes

Flow graph

Time-
proportional

Flow-
dependent

Event signal

Event-
proportional

A sample is extracted in fixed
time intervals (e.g. every 10
minutes) with a fixed sample
volume (e.g. 50 ml).

A sample is extracted in
variable time intervals
(depending on the flow (Q))
with a fixed sample volume.

The sampler is waiting for an
event (e.g. the exceeding of
a pH limit value). A sample is
extracted in fixed time
intervals (e.g. every 10
minutes) with a fixed volume
(e.g. 50 ml) as long as an
event is present.