Operation – Hach-Lange BUHLER 2000 Water sampler User Manual

Page 24

Advertising
background image

Operation

Page 24

Manual BÜHLER 2000/1029

6.2.1

**********************

FILLING TIME PER
BOTTLE

2:00H

**********************

TIME-RELATED bottle filling:
Enter for how long each bottle shall be filled until the distributor
changes to the next bottle.
Example:

Interval:

10 min.

Filling time/bottle: 120 min.
Bottle volume:

2,5 L

That means 12 samples are extracted in 120 min. With a bottle
volume of 2,5 L, the maximum dosing volume per sample is
calculated as follows:
2500:12

200 ml.

5.2.5

**********************

BOTTLE VERSION
*SINGLE BOTTLE
2 x 5,0 L
2 x 10,0 L

**********************

If your sampler is equipped with a distributor, select here the
distributor/bottle variant.

5.2.12

**********************

REPEAT PROGRAM

YES
*NO

**********************

If the program shall only run once, select NO.

Selecting „YES“ means continuous operation (the program is
always restarted).

Remark:

In that case, the program will run endlessly. So, please
make sure that the bottles are emptied regularly
according to the programming as otherwise they will be
overfilled.

1.1.4

**********************

PROGRAM NO. 1
TIME
INTERVAL HRS. 0:06
FILLING TIME H 2:00

**********************

Each programming is completed with this menu to have the pos-
sibility to check the program settings of the just created program.

Program No. 1 has been programmed with a time-proportional
program
.

To start the program, press the „DOWN“ arrow key.

1.1.5

**********************

PROGRAM NO. 1
CHANGE
DELAYED START
*START

**********************

The just created program can either be
started immediately
started with a delay or
• changed