Operation – Hach-Lange BUHLER 2000 Water sampler User Manual

Page 32

Advertising
background image

Operation

Page 32

Manual BÜHLER 2000/1029

3.4.1

**********************

MODE
*EVENT
TIME
FLOW ANAL. 0-20 mA

**********************

EVENT-related sampling

By pressing the ENT key, the event-related sampling mode is
selected.

4.12.1

**********************

INTERVAL

00:10

**********************

Enter the sampling interval in minutes for sampling modes „TIME
or „EVENT“.

A sample is e.g. extracted every 10 minutes.

(If you edit a time or event program, the next menu is „BOTTLE
FILLING“)

3.4.1

**********************

MODE
TIME
*FLOW ANAL. 0-20 mA
FLOW ANAL. 4-20 mA

**********************

FLOW ANALOGUE

By pressing the ENT key, the sampling mode „FLOW ANALOGUE“
(analogue signal 0/4 – 20 mA) is selected.

4.10.1

**********************

SAMPLES PER HOUR
AT 20 mA

40

**********************

If „ANALOGUE“ (0/4-20mA) has been selected, enter here how
often samples shall be extracted at a maximum analogue flow
signal.
Example:
- Flow analogue 0-20 mA
- At 20 mA 40 samples/hour shall be extracted.
- If there are on an average 10 mA for a period of 1 hour, 20

samples will be extracted.

3.4.1

**********************

MODE
FLOW ANAL. 0-20 mA
FLOW ANAL. 4-20 mA
*FLOW DIGITAL

**********************

FLOW DIGITAL

By pressing the ENT key, the sampling mode „FLOW DIGITAL“ is
chosen.

4.11.1

**********************

PULSE DIVIDER

25

**********************

Enter after how many input pulses (potential-free make contacts) a
sample shall be extracted in the sampling mode flow „DIGITAL.

Example:
At a max. flow 1000 pluses/hour are delivered.

40 samples per hour shall be extracted. Thus the pulse divider
has to be set to 25.

5.2.2

**********************

BOTTLE FILLING

*TIME RELATED
SAMPLE RELATED

**********************

Select whether the bottle filling shall be time-dependent or whether
it shall depend on a pre-programmed number of samples.