Operation – Hach-Lange BUHLER 2000 Water sampler User Manual
Page 33
Operation
Page 33
Manual BÜHLER 2000/1029
6.2.1
**********************
FILLING TIME PER
BOTTLE
2:00H
**********************
TIME-related bottle filling
Enter for how long each bottle shall be filled until the distributor
changes to the next bottle.
Example:
Interval:
10 min.
Filling time/bottle:
120 min.
Bottle volume:
2,5 L
That means 12 samples are extracted in 120 min. With a bottle
volume of 2,5 L, the max. dosing volume per sample is
calculated as follows: 2500:12
200 ml.
5.2.3
**********************
NO. OF SAMPLES PER
BOTTLE LIMITED?
*YES
NO
**********************
TIME-related bottle filling
In addition to the entry of a bottle filling time you can also limit the
number of samples per bottle.
By means of this menu point a protection against overfilling at flow-
dependent sample extractions is guaranteed.
6.5.1
**********************
MAX. NO. OF SAMPLES
PER BOTTLE
5
**********************
TIME-related bottle filling
Enter the max. number of samples per bottle.
6.3.1
**********************
SAMPLES PER BOTTLE
1
**********************
SAMPLE-related bottle filling
If a sample-related bottle filling has been selected, enter here the
number of samples per bottle.
5.2.4
**********************
FILLING TIME PER
BOTTLE LIMITED?
*YES
NO
**********************
SAMPLE-related bottle filling
In addition to the entry of a „MAX. NO. OF SAMPLES PER
BOTTLE“, you can also limit the „FILLING TIME PER BOTTLE“.
That means, after expiration of the entered time the distributor
changes to the next bottle.
6.6.1
**********************
MAX. FILLING TIME
PER BOTTLE
2:00H
**********************
SAMPLE-related bottle filling
Enter the max. filling time per bottle.
5.2.5
**********************
BOTTLE VERSION
12
x 1,0 L
*12 x 2,9 L
24
x 0,4 L
**********************
Select your distributor/bottle variant.