Operation – Hach-Lange BUHLER 2000 Water sampler User Manual

Page 26

Advertising
background image

Operation

Page 26

Manual BÜHLER 2000/1029

5.2.2

**********************

BOTTLE FILLING

*TIME-RELATED
SAMPLE-RELATED

**********************

Select whether the bottle filling shall be

1. time-dependent or
2. depend on a pre-programmed number of samples.

6.2.1

**********************

FILLING TIME PER
BOTTLE

2:00H

**********************

At TIME-RELATED bottle filling:
Enter for how long each bottle shall be filled until the distributor
changes to the next bottle.
Example:
Interval:

10 min.

Filling time/bottle:

120 min.

Bottle volume:

2,5 L

That means 12 samples are extracted in 120 min. With a bottle
volume of 2,5 L, the maximum dosing volume per sample is
calculated as follows: 2500:12

200 ml.

5.2.3

**********************

NO. OF SAMPLES PER
BOTTLE LIMITED?
*YES
NO

**********************

In addition to the entry of a bottle filling time you can also limit the
number of samples per bottle.

6.5.1

**********************

MAX. NO. OF SAMPLES
PER BOTTLE

5

**********************

Enter the maximum number of samples per bottle.

5.2.5

**********************

BOTTLE VERSION
*SINGLE BOTTLE
2 x 5,0 L
2 x 10,0 L

**********************

If your sampler is equipped with a distributor, select here the
distributor/bottle variant.

5.2.12

**********************

REPEAT PROGRAM?

YES
*NO

**********************

If the program shall only run once, select NO.

Selecting „YES“ means continuous operation (the program is
always restarted).

Remark:

In that case, the program will run endlessly. So,
please make sure that the bottles are emptied
regularly according to the programming as
otherwise they will be overfilled.