Operation – Hach-Lange BUHLER 2000 Water sampler User Manual
Page 26
Operation
Page 26
Manual BÜHLER 2000/1029
5.2.2
**********************
BOTTLE FILLING
*TIME-RELATED
SAMPLE-RELATED
**********************
Select whether the bottle filling shall be
1. time-dependent or
2. depend on a pre-programmed number of samples.
6.2.1
**********************
FILLING TIME PER
BOTTLE
2:00H
**********************
At TIME-RELATED bottle filling:
Enter for how long each bottle shall be filled until the distributor
changes to the next bottle.
Example:
Interval:
10 min.
Filling time/bottle:
120 min.
Bottle volume:
2,5 L
That means 12 samples are extracted in 120 min. With a bottle
volume of 2,5 L, the maximum dosing volume per sample is
calculated as follows: 2500:12
200 ml.
5.2.3
**********************
NO. OF SAMPLES PER
BOTTLE LIMITED?
*YES
NO
**********************
In addition to the entry of a bottle filling time you can also limit the
number of samples per bottle.
6.5.1
**********************
MAX. NO. OF SAMPLES
PER BOTTLE
5
**********************
Enter the maximum number of samples per bottle.
5.2.5
**********************
BOTTLE VERSION
*SINGLE BOTTLE
2 x 5,0 L
2 x 10,0 L
**********************
If your sampler is equipped with a distributor, select here the
distributor/bottle variant.
5.2.12
**********************
REPEAT PROGRAM?
YES
*NO
**********************
If the program shall only run once, select NO.
Selecting „YES“ means continuous operation (the program is
always restarted).
Remark:
In that case, the program will run endlessly. So,
please make sure that the bottles are emptied
regularly according to the programming as
otherwise they will be overfilled.