Operation – Hach-Lange BUHLER 2000 Water sampler User Manual

Page 50

Advertising
background image

Operation

Page 50

Manual Bühler 2000/1029

APPENDIX A - Special programming

EXTENDED PROGRAM SETTINGS (displayed when programming)

5.2.1

**********************

BOTTLE FILLING
*COMB. TIME/SAMPLE
TIME-RELATED
SAMPLE-RELATED

**********************

Only at flow analogue/digital mode!

Select the COMBINED MODE „TIME/SAMPLE
RELATED“ by pressing the ENT key. Press enter again to input a
number of samples per bottle and a bottle filling time (press ENT
again). The bottle change is depending on which setting (time or
number of samples) is reached first.

6.2.2

**********************

ADJUST TIME FOR
CHANGE TO BOTTLE 1?
*YES
NO

**********************

ADJUST TIME FOR CHANGE TO BOTTLE 1

With this function a fixed allocation between bottle number and
time of the day is achieved!
The program start is possible at any time, but with the condition
that the sampler changes into a 24-hour cycle with bottle No. 1 at
a desired time.
Contrary to the delayed program start, the sample extraction is
started immediately after the program start.

Remark:
This function is recommended if a program repeat has been
programmed as the change to bottle No. 1 after the program start is
always effected at the entered time in a 24-hour rhythm. If the
program has been started without program repeat, the bottle
change is effected as described and the program is terminated after
12 bottles have been filled.

A detailed description (example) can be found on the next page.

7.1.1

**********************

START TIME FIRST
BOTTLE

00:00

**********************

Enter the time at which the sampler shall change to bottle No. 1.