Operation – Hach-Lange BUHLER 2000 Water sampler User Manual

Page 45

Advertising
background image

Operation

Page 45

Manual Bühler 2000/1029

5.16.1

**********************

DIGITAL OUTPUT 1:
*PROGRAM ACTIVE
PROGRAM COMPLETED
BOTTLE CHANGE

**********************

If e.g. output 1 is selected, the opposite menu is displayed.
By pressing the ENT key, a list of possible signals is displayed.
Scroll with keys . Press ENT to mark the desired signal(s) with
a *. Pressing the ENT key again deletes the marking.

Remark:

See appendix for all selectable
messages.

4.17.28

**********************

DIGITAL INPUTS
*INPUT

1

INPUT

2

INPUT

3

**********************

Three digital inputs are available. Each of them can be selected and
configured.

5.17.1

**********************

DIGITAL INPUT 1:
*START PROGAM
STOP PROGRAM
SAMPLING

**********************

If e.g. input 1 is selected, the opposite menu is displayed.
By pressing the ENT key, a list of choices is shown. Scroll with keys
. Press the ENT key to mark the desired choice (action that
shall be triggered by an incoming signal) with a *. Pressing the ENT
key again deletes the marking.

4.17.31

**********************

BAUD RATE RS232
38400
57600
*115200

**********************

Here you can adjust the Baud Rate of the RS232 interface.

Remark:

In case of operation by modem, the baud rate set in this
menu has to be identical with the baud rate of the
modem used!

4.17.32

**********************

CHANGE PASSWORD

6299

**********************

ACTIVATION / DEACTIVATION OF PASSWORD
Here you can determine your own password. By pressing the ENT
key, a 4-digit code can be input. Press the ENT key again after the
entry. Now, the code is activated! If you want to deactivate it again,
input 0000 and confirm with ENT!

Remark:

In any case, note down the input code as otherwise
you will have no more access to the protected program
sections!

MEMORY

1.1.9

**********************

MEMORY

**********************

Here you can have a look at the stored information.

Remark:

By means of the optional RS232 interface kit you
can also read out the data to a computer.

Advertising