Operation – Hach-Lange BUHLER 2000 Water sampler User Manual

Page 41

Advertising
background image

Operation

Page 41

Manual Bühler 2000/1029

6.10.1

**********************

DELETE INFO MEMORY
AT PROGRAM START?
YES
*NO

**********************

Select whether the info memory shall be deleted at the program
start.

CHANGE SYSTEM SETTINGS

2.8.4

**********************

SYSTEM SETTINGS
CHANGE

**********************

Here you can change the system settings

Remark:

Modifications should only be carried out by specially
trained service personnel.
Wrong adjustments can lead to malfunctions or may
even damage the apparatus!

3.10.1

**********************

PASSWORD

****

**********************

This section is protected by a password. The factory-set password
is 6299 (the factory always activates the
password).

The password can also be changed.

DEACTIVATION:
To deactivate the password enter 0000 !

4.17.1

**********************

PRE-PURGE
PRIOR TO SAMPLING

5.0 s

**********************

Set the purging time prior to each sample extraction. This is
recommended in order to remove possible deposits in the suction
aperture.

4.17.2

**********************

POST-PURGE
AFTER SAMPLING

0.0 s

**********************

Set the purging time after each sample extraction (if the suction
hose is laid with a continuous fall, you can leave it at 000).
This purging is recommended if the suction hose is laid horizontally
in order to blow out remaining water.

4.17.3

**********************

AERATION TIME

5.0 s

**********************

Set the aeration time: time between filling level signal and opening
of pinch-valve in order to fill the sample into the bottle.

Advertising