Operation – Hach-Lange BUHLER 2000 Water sampler User Manual
Page 29
Operation
Page 29
Manual BÜHLER 2000/1029
6.2.1
**********************
FILLING TIME PER
BOTTLE
2:00H
**********************
At TIME-RELATED bottle filling:
Enter for how long each bottle shall be filled until the distributor
changes to the next bottle.
Example:
Interval:
10 min.
Filling time/bottle: 120 min.
Bottle volume:
2,5 L
That means 12 samples are extracted in 120 min. With a
bottle volume of 2,5 L, the maximum dosing volume per
sample is calculated as follows:
2500:12
200 ml.
6.3.1
**********************
SAMPLES PER BOTTLE
1
**********************
At SAMPLE-RELATED bottle filling:
Enter the maximum number of samples per bottle. As soon as the
set number of samples has been extracted, sampling into the
respective bottle is terminated.
5.2.5
**********************
BOTTLE VERSION
*SINGLE BOTTLE
2 x 5,0 L
2 x 10,0 L
**********************
If your sampler is equipped with a distributor, select here the
distributor/bottle variant.
5.2.12
**********************
REPEAT PROGRAM ?
YES
*NO
**********************
If the program shall only run once, select NO.
Selecting „YES“ means continuous operation (the program is
always restarted).
Remark:
In that case, the program will run endlessly. So,
please make sure that the bottles are emptied
regularly according to the programming as
otherwise they will be overfilled.
1.1.4
**********************
PROGRAMM NO. 3
EVENT
INTERVAL HRS 00:06
BTL CHANGE EVENT
**********************
Each programming is completed with this menu to have the
possibility to check the program settings of the just created
program.
Program No. 3 has now been programmed with an event-related
program.
To start the program, press the „DOWN“ arrow key.
1.1.5
**********************
PROGRAM NO. 3
CHANGE
DELAYED START
*START
**********************
The just created program can either be
• started immediately
• started with a delay or
• changed.