Operation – Hach-Lange BUHLER 2000 Water sampler User Manual

Page 49

Advertising
background image

Operation

Page 49

Manual Bühler 2000/1029

3.18.20

**********************

OUTPUTS
OUTPUT 1
OUTPUT 2
OUTPUT 3

**********************

Here you can check the outputs.

If you press ENT and a relay is connected, you can check whether
the output is o.k.

3.18.21

**********************

INPUT SIGNALS
FILLING LEVEL 100
FLOW ANAL

12.2 mA

EVENT 0 / FLOW DIG 0

**********************

The status of various inputs can be inquired.

Example:

FILLING LEVEL 250 (or higher)

Level electrodes are bridged (the respective value is directly re
lated to the conductivity. However, the indicated value has no
dimension)

• FLOW ANALOGUE 12 mA

There are 12 mA at the analogue input.

• EVENT 0 (1)

0 = There is no event signal
1 = There is an event signal

• FLOW (digital) 0 (1)

0 = No pulse
1 = Pulse (digital)