LabMaster 9 Zi-A

104

922157-00 Rev A

Sequence Mode Setup

When setting up Sequence Mode, you define the number of fixed-size
segments acquired in single-shot mode (see the instrument specifications
for the limits). The oscilloscope uses the sequence timebase setting to
determine the capture duration of each segment. Along with this setting,
the oscilloscope uses the number of segments, maximum segment length,
and total available memory to determine the actual number of samples or
segments, and time or points.

Setting up Sequence Mode (Adjacent)

1. Touch Timebase → Horizontal Setup... on the menu bar.

2. Click the Sequence tab.

3. Under Acquisition Settings, touch inside the Num Segments data

entry control and provide a value using your preferred input control
method. Additional information on using the touch screen controls
can be found in the Dialog Area (on page 87).

NOTE:

The number of segments you choose to display can be less than

the total number of segments in the waveform. For example, in the pop-
up images above, the number of display segments is 10, but the total
number of segments entered in the timebase dialog's Num Segments
control is 100.