Gently pull the green tab on the cable plug to remove it.

Repeat the previous steps for every additional Acquisition Module in your
system.

PLEASE NOTE THE FOLLOWING:

PCIe 4 Lane Cable plugs are keyed with a single groove along one wide
side of the plug. The plug must be inserted into the socket with the
groove aligned properly.

Be sure to connect your Acquisition Modules to the correct channel
groupings on the Master Control Module; First Acquisition Module is
connected into DATALINK CHANNEL 1-4, second into the DATALINK
CHANNEL 5-9 input, third into the DATALINK CHANNEL 9-12 input,
fourth into the DATALINK CHANNEL 13-16 input, and last into the
DATALINK CHANNEL 17-20 input.

If you are connecting less than 5 of the Slave Acquisition Modules, the
PCIe 4 Lane Data Inputs on the back of the Master Control Module may
not be skipped and must be connected in consecutive order from the PCIe
4 Lane Data Outputs on the back of the corresponding slaves.