Teledyne LeCroy LabMaster 9Zi-A User Manual
Page 133

Getting Started Manual
922157-00 Rev A
125
1.
From the menu bar, choose Vertical > Channel <#> Setup. Touch the
ChannelSync Cal tab to the right of the C<#> tab.
2.
Place a checkmark next to the module(s) you want to calibrate and
touch the ChannelSync Cal button. The ChannelSync Calibration
Wizard opens.
3.
Select to measure Input A or Input B, or select Active Input to make
measurements on the active input of each channel.
4.
Select a Ref Channel to be used as a reference against which all
other measurements are made during your calibration.
5.
Enter the V/div (gain) setting for which you want to perform the
ChannelSync Cal on all channels. The default value is 100 mV/div,
which corresponds to the Fast Edge output amplitude.
NOTE: Connect the Fast Edge calibration signal (located on the front of
the Acquisition Module) to the reference input channel (in this example,
C1), then touch the Next button. The ChannelSync™ Cal begins.