Labmaster 9 zi-a – Teledyne LeCroy LabMaster 9Zi-A User Manual
Page 150

LabMaster 9 Zi-A
142
922157-00 Rev A
Training TriggerScan
The TriggerScan Trainer inspects a currently acquired waveform and
automatically builds a list of common trigger setups used to find rare
events. To train TriggerScan:
Acquire and display at least three cycles of a signal before running
the Trainer.
Re-run the Trainer if you want to change the trigger types or if you
change the channel or signal.
1. Touch Trigger → Trigger Setup... from the menu bar.
2. On the Trigger dialog, click the TriggerScan tab.
3. Touch the Trainer button and select a channel as the source for the
training on the Trigger Scan Trainer pop-up. Also choose the types
of triggers the Trainer should use to train the system and then touch
the Start Training button. The training begins.
When it is complete, a list of smart trigger setups is displayed in the
Trigger List.