Getting Started Manual

922157-00 Rev A

35

Cap off any unused SMA 10 GHz Clock Outputs on the back of your 9CZi-A
Master Control Module using the chain-linked 50 Ω terminations provided.

Power Cable Connections and Main Power Switch

The combined draw from your 9CZi-A Master Control Module and 9xxSZi-A
Slave Acquisition Modules (max of 5) approximately totals 5 kW, Teledyne
LeCroy advises that your power connections be generally divided as
follows:

The Master Control Module requires less power and may be
combined with a Slave Acquisition Module to a different circuit
rated to handle a combined 2050 Watts.

Connect any additional pairs of Slave Acquisition Modules to a
different circuit rated to handle a combined 2050 Watts.

A 15A/250V 14AWG rated grounded cord set with a IEC320 right-angle
Type C15 connector with a slotted groove is provided specifically for the
Master Control Module. Components other than the Master Control
Module
are each provided with a standard 10A/250V 18AWG rated
grounded cord set with an IEC320 right-angle Type C13 connector. Ω

NOTE: Always refer to the product datasheet specifications at teledynelecroy.com
for the most current and detailed power ratings.