Maximize the application by touching the shortcut icon located on the
lower-right of the screen.

A sequence of calibration screens are shown when the Touch-Screen
Calibration
button is touched. These screens instruct you to touch a series
of displayed crosses in order to improve precision and accuracy when using
your touch screen.

The Service button is a section of the application reserved for qualified
Teledyne LeCroy personnel. When touched, an access code pop-up is
shown enabling Teledyne LeCroy technicians access to various service
dialogs inside the oscilloscope software.

PLEASE NOTE THE FOLLOWING:

Use a stylus for this procedure. Using your finger for the calibration
prompts provides insufficient accuracy for the procedure.

The calibration procedure times out after ten seconds pass and no cross is
touched.

Avoid parallax errors by placing your line of sight directly in front of each
cross before touching it.