Power Cable Connections and Main Power Switch

Teledyne LeCroy advises that you note the power ratings of the individual
modules and connect them to suitably rated circuits.

MCM-Zi Control Module comes with an IEC60320 Type C13 power
inlet connector.

9xxSZi-A Acquisition Module comes with an IEC60320 Type C20

power inlet connector.

Always refer to the product datasheet at www.teledynelecroy.com for the
most current and detailed specifications regarding power ratings.

CAUTION

: The combined draw from your Master Control Module and

Acquisition Modules (max of 20) can be many kilowatts. Two acquisition

modules connected to a single circuit may cause an overcurrent condition for that
circuit.

When all connections are made, the main power switch on the front of the
Master Control Module
powers on all components as a single unit.

NOTE

: If any connections are incorrect, the main power switch will not power on

the system.