Labmaster 9 zi-a – Teledyne LeCroy LabMaster 9Zi-A User Manual
Page 138

LabMaster 9 Zi-A
130
922157-00 Rev A
G
LITCH
Glitch trigger is a simpler form of Width trigger. Use Glitch trigger when
you want to define a fixed pulse-width time or time range only. Glitch
trigger makes no provision for voltage levels or ranges.
D
ROPOUT
Used primarily in single-shot applications, and usually with a pre-trigger
delay, Dropout trigger can detect lost signals. The trigger is generated at
the end of the timeout period following the last trigger source transition.
R
UNT
The Runt trigger occurs when a pulse crosses a first threshold line, but fails
to cross a second threshold line before re-crossing the first. Other defining
conditions for this trigger are the edge (triggers on the slope opposite to
that selected) and runt width.
S
LEW
R
ATE
SlewRate trigger activates a trigger when the rising or falling edge of a
pulse crosses two threshold levels: an upper level and a lower level. The
pulse edge must cross the thresholds faster or slower than a selected
period of time.