Power Cable Connection and Main Power Switch
The combined draw from your 9xxMZi-AMaster Acquisition Module,
9xxSZi-ASlave Acquisition Modules (max of 4), and CPU Module
approximately totals 5 kW, Teledyne LeCroy advises that your power
connections be generally divided as follows:
Connect the Master Acquisition Module and its CPU Module to a
single circuit rated to handle a combined 1250 Watts.
Connect any additional pairs of Slave Acquisition Modules to a
different circuit rated to handle a combined 2050 Watts.
NOTE: Always refer to the product datasheet specifications at teledynelecroy.com
for the most current and detailed power ratings.
15A/250V 14AWG rated grounded cord set with a IEC320 right-angle Type
C15 connector with a slotted groove is provided specifically for the Master
Acquisition Module. Components other than the Master Acquisition
Module are each provided with a standard 10A/250V 18AWG rated
grounded cord set with an IEC320 right-angle Type C13 connector.