Passive probe compensation – Teledyne LeCroy LabMaster 9Zi-A User Manual
Page 79
Advertising
![background image](/manuals/353442/79/background.png)
Getting Started Manual
922157-00 Rev A
71
Passive Probe Compensation
Passive probes must be compensated to flatten overshoot. This is
accomplished by means of a trimmer at the connector end of the probe.
1. Attach the connector end of your passive probe to any channel.
2. Connect the probe end to the CAL output connector at the front of
the oscilloscope. Ground the probe.
3. Adjust the trim pot at the connector end of the probe until the
square wave is as flat as possible.