REV 1 TRM-203 AND TRM-403 USER’S MANUAL

39

The TRM will then ramp up the current and the screen will be updated accordingly:

The TRM determines when the resistance reading is stable and displays the resistance
value on the LCD screen as shown below:

Press the

[ENTER]

key to save the reading.

l. The following screen will be displayed:

After the reading is saved, the TRM will continue to test the resistance and display the
readings. You may press the

[ENTER]

key again to save the current reading. Press the

[STOP]

key when done.

m. The following screen will be displayed:

Press the

[1]

key (YES) to print the test results or the

[2]

key (NO) if you do not want to

print the test results. A sample single phase resistance test printout is shown in Figure 6.

PRINT TEST RESULTS?
1.

YES

2.

NO




*XFMR CHARGING*


I = 39.91 AMPS

R = 1.491

m

Ω

===

>

TEST SAVED!

<

===

** XFMR ENERGIZED! **

TEST IN PROGRESS 7

I = 39.91 AMPS

R = 1.491

m

Ω



*XFMR CHARGING*

H/X1-H/X0

please wait...

I = 39.98 AMPS

** XFMR ENERGIZED! **

Advertising