<8. VALIDATION AND CALIBRATION>

MANUAL OffLINE VALIDATIONS

Enter into the Advanced Menu, Calibrate &
Validate section, Off-Line Validations.
Choose Check gas 1, 2 or two gas validation
and select Manual Validation.

Following the on screen directions, enter
in the pressure, temperature, cell length, &
concentration of the gas within the off-line line
check gas flow cell. Press Enter to proceed.