Basic operation – Yokogawa In-Situ Gas Analyzer TDLS200 User Manual

Page 60

Advertising
background image

6-3

6-4

IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00

<6. BASIC OPERATION>

Display Text

Description

Line 1 - Measurement

O2 xx.x %

Measured gas and unit of measurement For Gas 1

and/or Gas 2 as configured

moisture xx.x ppm

Line 2 – Transmission or
Second Gas Measurement

Transmission xx.x %

Laser light transmission strength

(0-100% range)

CH4 xx.x %

Second measurement gas and unit

Line 3 - Status

Initializing……

shown during the power-up and

initialization of the analyzer

System OK

Normal Operation condition with no active alarms

WARNING Det Sig Low

WARNING Trans Low

WARNING Spectr Noise

WARNING Gas Pres

WARNING Gas Temp

WARNING Gas Level

WARNING Board Temp

WARNING Conditions

FAULT Laser Temp

FAULT Det Sig High

FAULT Det Sig Lost

FAULT Peak Response

FAULT Peak Center

FAULT Conditions

Zero Calibrating…

Span Calibrating…
Offline Validating…
Online Validating…

Validation Status

Data Transferring…

Transfer Success

Transfer Failure

Data Transfer Status

Line 4 - Information

Yokogawa TDLS

SN 76-1xxx-05-xx

AO1: CONC xx-xx%/ppm/ppb/mg/m3

or mg/Nm3

AO2: TRANS xx-xx%

AO3: TEMP xx-xxF/C/K

10.0.0.35

TEMP Act/Con/Fox xx F/C/K

PRES Act/Con/Fix xx.x PsiA/BarA/KPa/

torr/atm

OPL xx.x in/ft/cm/m

Launch xx deg C

Detect xx deg C

Analyzer Name

Analyzer Serial No.

Configured 4-20mA output for AO1, AO2, & AO3

Static IP Address

Process Gas Temperature used for gas

concentration calculation

Process Gas Pressure used for gas

concentration calculation

Optical Path Length over which the analyzer is

measuring the target gas

Launch unit internal temperature

Detect unit internal temperature

Online Menu

Level 1 Menu Level 2 Menu

Level 3 Menu

Level 4 Menu

Level 5 Menu

Trends

Refresh

Refresh Current Trend screen

Gas 1 Concentration

STDEV of Gas 1 Concentra-
tion*
Gas 2 Concentration*
STEV of Gas 2 Concentration*
Transmission*
Laser Temp Setpoint*
Laser Temp in DegC*
Peak Center Position*
Gas Temperature*
Gas Pressure*

*(Similar to Gas 1 Concentration)