Following pages are the various “Laser Spectra & Control” screen captures that depict the different scenarios
depending on what process gas(es) are present in the OPL and what gas is purged through the analyzer:

Note that the actual magnitude of the absorption units (au) as indicated on the “Absorption Spectrum” Y-Scale
(vertical) will vary depending upon the actual gas concentrations, optical path lengths, gas temperatures and gas
pressures.

TDLS200 TDL Analyzer Instruction Manual V2.1

Page 73 of 131

-

3

7

-

Following pages are the various “Laser Spectra & Control” screen captures that depict the different

scenarios depending on what process gas(es) are present in the OPL and what gas is purged

through the analyzer:

Note that the actual magnitude of the absorption units (au) as indicated on the “Absorption

Spectrum” Y-Scale (vertical) will vary depending upon the actual gas concentrations, optical path

lengths, gas temperatures and gas pressures.

Off-Line Calibration Conditions:

Process/Calibration OPL: N

2

Analyzer Purge: N

2

This is how the absorption spectra will appear when

there is neither measured or reference gas (i.e.NH

3

and H

2

O) in the entire optical path (Process OPL

and Analyzer Purge).

required for an Off-Line Zero

Calibration of the analyzer.

Process/Calibration OPL: ~20 ppm NH

3

Analyzer Purge: N

2

In this Absorption Spectra view, there is practically

zero reference gas (H

2

O) absorption peak in the

approximate peak center position (PCP) 167 region.

There is approximate 20 ppm of me a s ured gas

(NH

3

) (@ approx PCP ~120) absorption in this

spectra based on 72.6cm OPL, ambient

temperature and ambient pressure.

required for an Off-Line Span

Calibration of the analyzer.

This is the spectra appearance

This is the spectra appearance

Process/Calibration OPL: N2
Analyzer Purge: N2

This is how the absorption spectra will appear
when there is neither NH

3

nor H

2

O in the entire

optical path (Process OPL and Analyzer Purge).

This is the spectra appearance
required for an Off-Line Zero
Calibration of the analyzer.

Process/Calibration OPL: ~20ppm NH

3

Analyzer Purge: N

2

In this Absorption Spectra view, there is practi-
cally zero H

2

O absorption peak in the approxi-

mate peak center position (PCP) 167 region.

There is approximate 20 ppm NH

3

(@ approx

PCP ~120) absorption in this spectra based on
72.6cm OPL, ambient temperature and ambient
pressure.

This is the spectra appearance
required for an Off-Line Span
Calibration of the analyzer.