9 hazardous area systems – Yokogawa In-Situ Gas Analyzer TDLS200 User Manual

Page 53

Advertising
background image

<5. INSTALLATION AND WIRING>

5-19

5-20

IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00

5.11.1 Purging Analyzer for Hazardous Areas (with On-Line Validation)

• NEC/CSA Class 1, Division 2, Groups A-D

ATEX Zone 2 CAT 3 Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge control
system is fitted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pressure switch
alarm contacts (open on loss of purge pressure). The purging should be carried in sequence typically as shown
below. All purge gas connections are ¼” od Tube fittings.

5.11.2 Purging Analyzer for Hazardous Areas (without On-Line Validation)

• NEC/CSA Class 1, Division 2, Groups A-D
• ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge
control system is fitted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pres-
sure switch alarm contacts (open on loss of purge pressure).

Figure 48 - Purge Flow Diagram when not using on line validation

TDLS200 TDL Analyzer Instruction Manual V2.1

Page 49 of 131

- 49 -

3.9.2 Purging Analyzer for Hazardous Areas (without On-Line Validation)

Z-Purged designed in accordance with NEC/CSA Class 1, Division 2, Groups A-D

Certified ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging.

A Z-Type purge control system is fitted the Launch Unit and it includes a local indicator

(Bright Green, rugged light) and pressure switch alarm contacts (open on loss of purge

pressure).

Figure 28 - Purge Flow Diagram when not using on line validation

3.9.3 Purging Analyzer and Universal Power Supply and/or URD for Hazardous

Areas (with On-Line Validation)

Z-Purged designed in accordance with NEC/CSA Class 1, Division 2, Groups A-D

Certified ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging when

using in conjunction with either or the Universal Power Supply and Universal Remote Display. A Z-

Type purge control system is fitted the Launch Unit and it includes a local indicator (Bright Green,

rugged light) and pressure switch alarm contacts (open on loss of purge pressure).

Figure 29 - Purge Flow Diagram when using on line validation

Main

Electronic

Housing &

Purge

System

Laser

Module

Check Gas

Flow Cell

Proce

ss

Inte

rfa

ce

Proce

ss

Inte
rfa

ce

Detect

Module

Detect

Electronic

Housing

Nitrogen or I/A Purge

Gas

Nitrogen or I/A Purge

Gas

Power

Supply

or URD

Main

Electronic

Housing &

Purge

System

Laser

Module

Check Gas

Flow Cell

Pr

oce

ss

Int

erf

ac

e

Pr

oce
ss

Int

erf

ac

e

Detect

Module

Detect

Electronic

Housing

Nitrogen or I/A Purge Gas

Main

Electric

Housing

& Purge

System

Detect

Electronic

Housing

TDLS200 TDL Analyzer Instruction Manual V2.1

Page 48 of 131

- 48 -

3.9 Hazardous Area Systems

The TDLS200 Analyzer requires a continuous nitrogen gas purge to prevent ambient oxygen

ingress to the optical path, when oxygen is the measured gas. The flow rate can be minimized as

long as it prevents any ambient oxygen ingress to the measurement optical path. Other purge

gases may be used as long as they do not contain any of the measured gas and are clean, dry,

etc.
For hazardous area operation, the same nitrogen purge gas is used to purge the entire analyzer

(including non-optical path sections such as the electronics).
The process interface may also require purging to maintain clear windows, refer to Process

Window Purge details separately.

NOTE: Please also refer to any separate Purge System Original Manufacturers Operating

Instructions and Manuals in conjunction with this Instruction Manual.

The Purge Systems are not manufactured by Yokogawa Laser Analysis Division.

3.9.1 Purging Analyzer for Hazardous Areas (with On-Line Validation)

Z-Purged designed in accordance with NEC/CSA Class 1, Division 2, Groups A-D

Certified ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!

The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-

Type purge control system is fitted the Launch Unit and it includes a local indicator (Bright Green,

rugged light) and pressure switch alarm contacts (open on loss of purge pressure). The purging

should be carried in sequence typically as shown below. All purge gas connections are ¼” od Tube

fittings.

Figure 27 - Purge Flow Diagram when using on line validation

Main

Electronic

Housing &

Purge

System

Laser

Module

Check Gas

Flow Cell

Proce

ss

Inte

rfa

ce

Proce

ss

Inte
rfa

ce

Detect

Module

Detect

Electronic

Housing

Nitrogen or I/A

Purge Gas

Nitrogen or I/A

Purge Gas

Figure 47 - Purge Flow Diagram when using on line validation

Main

Electric

Housing

& Purge

System

Detect

Electronic

Housing

Nitrogen or I/A Purge
Gas

Check Gas

Flow Cell