Installation and wiring> 5-16 – Yokogawa In-Situ Gas Analyzer TDLS200 User Manual

Page 49

Advertising
background image

<5. INSTALLATION AND WIRING> 5-16

IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00

5.10.2 Launch and Detect Unit Wiring – Standard for CE/ATEX

Launch Unit (to Detect) Terminations TB-7 (ensure 360˚ cable shield ground)

Detect Unit (to Launch) Terminations TB-7 (ensure 360˚ cable shield ground)

SPECIAL NOTE: Noise sources magnetically coupled to the 4-pair launch to detect cable can cause errors to
the measurement if the frequency and applied total voltage meet or exceed the following; Frequency range of
0.1 to 7.5 MHz and Total induced voltage of 3Vrms.

A combination of these parameters will cause measurement error outside of the analyzer performance
specification. In practice, these potential errors can be easily avoided by routing the 4-pair launch to detect
cable away from electromagnetic interference sources that could exhibit these parameters. Such sources might
include power transformers, electric motors, electric welding machinery, high voltage power lines, etc. Analyzer
grounding wires and any other I/O lines such as 8-pair launch to utility panel cable or A I/O cables to/from the
DCS should also be installed with similar basic practices to also ensure there is no additional adverse influence
on any of the I/O signals.

 

 

 