Analyzer components – Yokogawa In-Situ Gas Analyzer TDLS200 User Manual

Page 29

Advertising
background image

4-10

IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00

<4. ANALYZER COMPONENTS>

4.7 Analyzer Connections

Launch – Detect Interconnect
The two units are connected to each other via a four, twisted pair cable suitable for tray installation outdoors.
Pluggable terminals strips are provided at both units to enable field termination of the cable. The cable pairs are
individually shielded as well as an overall shield.

The cable specifications are as below.

Item

Specification

Number fo Pairs

4, individually shielded

Total Number of Conductors

9 (includes 1 comm.)

AWG

18 (0.75 mm

2

)

Conductors

7 x 26 stranding, Bare Copper

Inner Shield

Aluminum Foil-Polyester tape, 100% coverage with 20
AWG tinned copper wire drain

Insulation

F-R PVC – Flame Retarding Polyvinyl Chloride

Outer Shield

Aluminum Foil-Polyester tape, 100% coverage with 18
AWG tinned copper wire drain

Outer Jacket

F-R PVC – Flame Retarding Polyvinyl Chloride Wall
thickness 0.053” (1.35 mm) Typical 0.47” (12 mm)
outside diameter

Operating temperature

-22 to 221ºF (-30 to +105ºC)

Min. Bend Radius

5” (127 mm)

Applicable Standards

NEC/(UL) PLTC, ITC, CMG

Flame Test

UL1581, FT4, IEEE 1202 & ICEA T-29-520

Suitability

Indoor, Outdoor, Burial and Sunlight Resistant
Power Limited Tray Rated Cable

Nom. Conductor DC resistance @ 20˚C

5.86 Ohms/1000 ft (305 m)

Nom. Outer Shield DC resistance @ 20˚C

4.75 Ohms/1000 ft (305 m)

Max. Operating Voltage - UL

300 V RMS

Conductor Identification

Numbered pairs, black & white conductors

Typical Manufacturer & Part No.

Belden Type 1475 A

The maximum cable length should not exceed 150 ft (46 m).

Please ensure that the Launch to Detect cable is properly terminated and that all grounding and shielding
details are correct per installation drawings-Espically important for CE/A TEX
installations.