On-Line Process Conditions – Analyzer Purged with Nitrogen:
When purging with Instrument Air (as opposed to ambient air), the H

2

O levels are much smaller (due to it

having been dried and having a dew point typically in the order of -40˚C) and therefore the H

2

O

concentration is not large enough to produce an absorption peak that can be seen.

 

Process OPL: Zero NH

3

+ Zero H

2

O

Analyzer Purge: Air (typical ~2-4% H

2

O)

In this Absorption Spectra view, the analyzer air purge H

2

O

absorption peak can be seen at approximate peak center position
(PCP) 167.

There is no NH

3

absorption in the spectra.

 

Process OPL: ~20 ppm NH

3

+ Zero H

2

O

Analyzer Purge: Air (typical ~2-4% H

2

O)

In this Absorption Spectra view, the H

2

O absorption (analyzer air

purge) peak can be seen at approximate peak center position (PCP)
167.

There is approximate 20 ppm NH

3

absorption (approx 120PCP)

in this spectra based on 72.6 cm OPL, ambient temperature and
ambient pressure.

 

Process: Zero NH

3

+ ~2% H

2

O

Analyzer Purge: Air (typical ~2-4% H

2

O)

In this Absorption Spectra view, the process gas H

2

O (~2% 0.7 m

@ STP) and analyzer air purge H

2

O absorption peak can be seen at

approximate peak center position (PCP) 167.

There is no NH

3

absorption in the spectra.