Basic operation – Yokogawa In-Situ Gas Analyzer TDLS200 User Manual

Page 80

Advertising
background image

6-24

IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00

<6. BASIC OPERATION>

On-Line Process Conditions – Analyzer Purged with Nitrogen:

Process OPL: NH

3

and H

2

O

Analyzer Purge: N

2

In this Absorption Spectra view, the Process H

2

O

absorption peak can be seen at approximate
peak center position (PCP) 167. This absorption
peak is Process H

2

O as there is no reference gas

H

2

O in the N

2

analyzer purge gas.

There is also approximate 20 ppm NH

3

(@ approx PCP

120) absorption in this spectra based on 72.6 cm OPL,
ambient temperature and ambient pressure.

Process OPL: Zero NH

3

+ ~2% H

2

O Air

Analyzer Purge: N

2

In this Absorption Spectra view, the large process
H

2

O absorption peak can be seen at

approximate peak center position (PCP) 167.

There is no NH

3

absorption in the spectra .

Process OPL: ~20 ppm NH

3

+ Zero H

2

O

Analyzer Purge: N

2

In this Absorption Spectra view, there is practically
zero process H

2

O absorbtion peak in the approximate

peak center position (PCP) 167 region.

There is approximate 20 ppm NH

3

(@ approx PCP

~120) absorption in this spectra based on 72.6 cm
OPL, ambient temperature and ambient pressure.

TDLS200 TDL Analyzer Instruction Manual V2.1

Page 74 of 131

- 74 -

On-Line Process Conditions – Analyzer Purged with Nitrogen:

Process OPL: Measured and Reference gas

(NH3 and H2O)
Analyzer Purge: N2

In this Absorption Spectra view, the reference gas

(H2O) absorption peak can be seen at approximate

peak center position (PCP) 167. This absorption

peak is from the reference gas contained in the

process stream ( H2O) as there is no reference gas

(H2O) in the analyzer purge gas.

There is also approximate 20ppm NH3 (measured

gas) (@ approx PCP 120) absorption in this spectra

based on 72.6cm OPL, ambient temperature and

ambient pressure.

Process OPL: Zero Measured Gas (NH3) + ~2%

H2O (Reference Gas)
Analyzer Purge: N2

In this Absorption Spectra view, the large reference

gas (H2O) absorption peak can be seen at

approximate peak center position (PCP) 167.

There is no measured gas (NH3) absorption in the

process .

Process OPL: ~20ppm Measured Gas (NH3) +

Zero Reference Gas (H2O)
Analyzer Purge: N2

In this Absorption Spectra view, there is practically

zero reference gas contained in the process stream.

No absorption from H2O at the approximate peak

center position (PCP) 167 region.

There is approximate 20ppm measured gas (NH3)

(@ approx PCP ~120) absorption in this spectra

based on 72.6cm OPL, ambient temperature and

ambient pressure.

TDLS200 TDL Analyzer Instruction Manual V2.1

Page 74 of 131

- 74 -

On-Line Process Conditions – Analyzer Purged with Nitrogen:

Process OPL: Measured and Reference gas

(NH3 and H2O)
Analyzer Purge: N2

In this Absorption Spectra view, the reference gas

(H2O) absorption peak can be seen at approximate

peak center position (PCP) 167. This absorption

peak is from the reference gas contained in the

process stream ( H2O) as there is no reference gas

(H2O) in the analyzer purge gas.

There is also approximate 20ppm NH3 (measured

gas) (@ approx PCP 120) absorption in this spectra

based on 72.6cm OPL, ambient temperature and

ambient pressure.

Process OPL: Zero Measured Gas (NH3) + ~2%

H2O (Reference Gas)
Analyzer Purge: N2

In this Absorption Spectra view, the large reference

gas (H2O) absorption peak can be seen at

approximate peak center position (PCP) 167.

There is no measured gas (NH3) absorption in the

process .

Process OPL: ~20ppm Measured Gas (NH3) +

Zero Reference Gas (H2O)
Analyzer Purge: N2

In this Absorption Spectra view, there is practically

zero reference gas contained in the process stream.

No absorption from H2O at the approximate peak

center position (PCP) 167 region.

There is approximate 20ppm measured gas (NH3)

(@ approx PCP ~120) absorption in this spectra

based on 72.6cm OPL, ambient temperature and

ambient pressure.

TDLS200 TDL Analyzer Instruction Manual V2.1

Page 74 of 131

- 74 -

On-Line Process Conditions – Analyzer Purged with Nitrogen:

Process OPL: Measured and Reference gas

(NH3 and H2O)
Analyzer Purge: N2

In this Absorption Spectra view, the reference gas

(H2O) absorption peak can be seen at approximate

peak center position (PCP) 167. This absorption

peak is from the reference gas contained in the

process stream ( H2O) as there is no reference gas

(H2O) in the analyzer purge gas.

There is also approximate 20ppm NH3 (measured

gas) (@ approx PCP 120) absorption in this spectra

based on 72.6cm OPL, ambient temperature and

ambient pressure.

Process OPL: Zero Measured Gas (NH3) + ~2%

H2O (Reference Gas)
Analyzer Purge: N2

In this Absorption Spectra view, the large reference

gas (H2O) absorption peak can be seen at

approximate peak center position (PCP) 167.

There is no measured gas (NH3) absorption in the

process .

Process OPL: ~20ppm Measured Gas (NH3) +

Zero Reference Gas (H2O)
Analyzer Purge: N2

In this Absorption Spectra view, there is practically

zero reference gas contained in the process stream.

No absorption from H2O at the approximate peak

center position (PCP) 167 region.

There is approximate 20ppm measured gas (NH3)

(@ approx PCP ~120) absorption in this spectra

based on 72.6cm OPL, ambient temperature and

ambient pressure.

Advertising