Sbc ethernet port, Basic operation – Yokogawa In-Situ Gas Analyzer TDLS200 User Manual

Page 91

Advertising
background image

6-35

IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00

<6. BASIC OPERATION>

6.9.1 Instructions for Connecting an External Computer to the Analyzer

Requires Windows 98SE or later (on computer to be connected to the analyzer) and crossover Ethernet cable.
The analyzer is provided with a CD that includes the VNCviewer program. Windows 7 will support the Ultra-
VNC software and associated functionality. Contact your local Yokogawa agent for a free copy of the necessary
“VNCviewer.exe” file that will enable the VNC connection with the analyzer. This VNCviewer.exe file should be
loaded on to the connecting PCs desktop ready for use when connecting with the analyzer.

• From Control Panel – Network Connections, make sure Ethernet Local Area Connection is set to
Enabled status. Disable wireless and any other networking connections.
• Connect crossover Ethernet cable from client system to Analyzer.
• On the computer, go to Control Panel – Network Connections - Local Area Connection, Internet
Protocol (TCP/IP) Properties.

o Set IP to 10.255.255.254 & Subnet Mask to 255.0.0.0. Select OK to accept changes on Internet

Protocol (TCP/IP) and Local Area Connection Properties.

o Start Ultra-VNC software, running from Desktop using the Guide below:

Figure 55 - Connecting External Computer to the analyzer

 

SBC Ethernet Port