If the windows are contaminated they may be cleaned using the following procedure:

Remove analyzer from process (isolate, etc. if necessary)

Use a clean, dry instrument air or nitrogen pressure supply to first blow off any particulate

matter.

Using warm water and mild soap detergent, gently clean the window surface with a soft,

non-abrasive cloth.

If the deposits do not come off then use a small amount of IPA (Isopropyl Alcohol) and a

soft, non-abrasive cloth.

Use the same clean, dry instrument air or nitrogen pressure supply to blow dry the surface.

Carefully check the entire surface of the window from different angles to ensure it is

thoroughly cleaned and ready for service.

If the window does not appear to clean up well, then replace the window assembly with a

new one.

If the window appears to have an etched surface then it has probably been contaminated

with HF or other similar corrosive gas.

Figure 32 - Window Cleaning

Figure 56 - Window Replacement

TDLS200 TDL Analyzer Instruction Manual V2.1

Page 89 of 131

- 89 -

Figure 33 - Window Replacement

5.1.2 Replacing Windows:

If it has been determined that a window required replacement, then the following should be

performed:

Shut down the analyzer and remove power, thus ensuring the laser if OFF.

Remove Launch or Detect (as appropriate) from the process interface by removing the one

fixed hole fastener (5/32” Allen Wrench), loosening the remaining three fasteners and then

twisting and pull off the unit.

Remove the four retaining screws (3/32” Allen Wrench, T-Handle) and lock washers from

the window holder recessed inside the quick connect body.

Remove the window that requires replacement.

Ensure the new replacement window is of the same specification as the window being

removed. Standard Windows are 0.25” Thick wedged, BK-7 borosilicate crown glass

mounted in stainless steel retaining ring.

The window holder assemblies can only be installed in one way by virtue of the mounting

hole placement.

Ensure a new O-Ring (of same specification) is installed at the same time. Viton is the

standard material however, certain applications may require other materials – check original

specifications.

With the new O-Ring in position, carefully locate the new window holder in place with the

smooth glass surface against the O-Ring.

Carefully tighten all four screws (ensuring the lock washers are in place) using a T-handle

wrench. Keep tightening all the fasteners in turn by ¼ turns thus ensuring an even torque

loading.

Hand tighten the screws until they are very firmly tightened (a T handle wrench as shown

below will torque its own shaft by about ¼ turn when the screws are fastened securely).

Offset screw position

for window

O-ring correctly

seated in position

O- ring correctly
seated in position