<9. TROUBLE SHOOTING>

9.1 Common Troubleshooting Steps

For most conditions the troubleshooting steps are common. In general, the most common issues
with the analyzer revolve around ensuring an adequate amount of the laser light is received at the
detector.

Check Status LEDs. This will ensure that power is routed properly to the system

components. Status LEDs are listed below:

Blue LED #

(from left to right)

Voltage VDC

Use

D20

+5

Laser temp control power supply

D26

-15

Isolated power supply

D27

+15

Isolated power supply

D24

+8

DFB laser driver power supply

D23

-8

DFB laser driver power supply

D22

-12

DFB laser driver power supply

D16

-12

Detector board power supply

D19

-12

VCSEL laser driver power supply

D6

-15

Analog I/O board power
supply

D12

-15

FPGA board power supply

D18

+12

VCSEL laser driver power supply

D7

+12

LCD power supply

D8

+24

Main power supply

D9

+15

Analog I/O board power
supply

D10

+5

SBC power supply

D11

+15

FPGA board power supply

D13

+5

FPGA board power supply

D14

+6

FPGA board power supply

D15

+12

Detector board power supply

D17

+12

General 12V power supply

 

 