MAIN MENU (Home Page)
Display of Concentration & Transmission (or 2

nd

Concentration)

Status Window – notification of initiating, working properly,
warnings or faults
Gas Temperature
Gas Pressure
Selection of Basic or Advanced Menu
Tag No and Serial No. (SN.) configured to analyzer
Active Alarm Display Button
Analyzer Shut Down Button

 

After selection of either Basic or Advanced Menu you will see
the Output Selection screen. PLEASE note the RIGHT ARROW
key access to the MENU (older version of software require no
other key or Press 9 to access MENU)
This allows control of the analog output while the user is
working in the analyzer software. Enter mA value form 0-20.
• Block will hold outputs at user selectable mA (example
shows 3.8mA) until return to Main Screen but

NOTE CHANNEL 2 ALWAYS TRACKS!

• Track will allow outputs to continue to report
concentration and transmission until return to Main
Screen
• Hold will hold outputs at their current value until return to
Main Screen

NOTE CHANNEL 2 ALWAYS TRACKS!

 

BASIC MENU
Configure – allows setting of Path Length, Gas
Temperature, Gas Pressure

View Spectra – user will select display of raw detector
signal or absorption spectra

Data – Alarm History, Calibration History

Trends – Allows for the displaying of data in a trend format

6.2 Software Guide

NOTE: At any time in the Main Menu, press the F5 key (only on Screen & Keypad versions)
to toggle the LCD backlight on/off. If there is no keypad (or VNC) activity for 30 minutes
then the LCD backlight will automatically switch off, press any key to restore backlight.