TDLS200 TDL Analyzer Instruction Manual V2.1

Page 19 of 131

- 19 -

2.3 Analyzer Components

Figure 1 - System Overview

Launch to Detect Interconnect (cable)

Launch Unit:

Main Electronics Housing

User Interface (optional)

Laser Assembly

Check Gas Flow Cell (for

On-Line Validation)

Detect Unit:

Detect Electronics

Housing

Detector

Assembly

Process Interface:

Analyzer detachable from process interface for Off-Line

calibration / service.

Flanged O-Ring Alignment

Flanged Metal Bellows Seal Alignment

Flow Cell

Isolation Flanges

By-Pass Piping

Custom designs for specific applications.

Hazardous Area Purge (optional)

NEC/CSA Class 1, Div. 2, Gr. A-D

ATEX Zone 2 Cat 3

Figure 4 - System Overview

4-1

IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00

<4. ANALYZER COMPONENTS>

Launch Unit:

• Main Electronics Housing
• User Interface (optional)
• Laser Assembly
• Check Gas Flow Cell (for
On-Line Validation)

Hazardous Area Purge (optional)

• NEC/CSA Class 1, Div. 2, Gr. A-D
• ATEX Zone 2 Cat 3

Launch to Detect Interconnect (cable)

Launch Unit:

• Analyzer detachable from process
interface for Off-Line calibration / service.
• Flanged O-Ring Alignment
• Flanged Metal Bellows Seal Alignment
• Flow Cell
• Isolation Flanges
• By-Pass Piping
• Custom designs for specific applications.

Detect Unit:

• Detect Electronics
Housing
• Detector Assembly

• The Launch Unit and Detect Unit are connected to each other via a Tray Rated 4-pair shielded twisted
pair cable.
• The Launch Unit requires a single 24VDC power supply (by customer or via optional Power Supply
Unit).
• Nitrogen purge gas is required to prevent ambient oxygen ingress however, for other target gases it
may be possible to use Instrument Air for purging.
• The Process Interfaces are available in various formats, sizes and materials to suit the desired
measurement/installation.
• The available Remote Interface Unit (RIU) can be located typically up to 100m (330ft) away from the
Launch Unit. The RIU also requires a 24VDC power supply. The RIU connects to the Launch Unit on
Ethernet (10-base-T 10/100) via CAT5e field rated cable.
• The available Universal Remote Display (URD) can be located typically up to 40m (120ft) away from the
Launch Unit. The URD requires an AC power input that is connected to a universal power supply with
24VDC output power supply (for the analyzer). The URD connects to the Launch Unit via multipair

sheilded twisted pair cable.

4 ANALYZER COMPONENTS