<4. ANALYZER COMPONENTS>

4.6 Process Interface
An appropriate Process Interface is selected to suit the process/stack installation. The analyzer is
detachable from the process interface to facilitate Off-Line calibration and service.

Process Interface Options
There are several systems available as well as custom designs for specific applications.

• Flanged O-Ring Alignment comprises typical 2” or 3”process
flange with a large diameter O-Ring seal, typically used for
stack or inert applications that are non-corrosive.
• Flanged Metal Bellows Seal comprises typical 2” or 3”
process flange with a metal bellows seal and external

mechanical alignment system, typically used when gas

containment is important.
• LAO-Large Aperature Optics, for long path combustion

application.
• Flow Cell may be used when the process gas has been

extracted or is used in a by-pass flow loop. This allows for

heat tracing (if necessary) and easy introduction of both Zero

and Span gases.
• Isolation Flanges are supplied with process windows

mounted in the flanges themselves typically for very

corrosive and/or high pressure applications.
• By-Pass Piping may be used when the process gas line is of

small diameter (typically <10”) and when no suitable section

of pipe work exists on this which the analyzer can be

mounted. By pass piping systems may be provided or

supplied at site. Heat tracing may also be supplied or

provided at site.
• Off Line Calibration Cell is used for off line verification of the

system. This cell is used to flow Zero and Span gases when

the analyzer is not connected to the process.

TDLS200 TDL Analyzer Instruction Manual V2.1

Page 29 of 131

- 29 -

The cable entry located on the bottom face of the enclosure. It is typically a ¾” Myers hub that has

a ¾” NPT female thread. It has a ground lug to facilitate the grounding of cable shields to the

analyzer chassis.

Detector Circuit Board

main function is to convert detector photocurrent into voltage and send it

to be sampled.
LEDs are incorporated to provide simple diagnostic of available power.
The board has a temperature sensing chip/circuit that monitors the ambient temperature inside the

detect electronics enclosure. The sensor is located on the top edge of the detect board to ensure

the maximum temperature reading is monitored.
The board is medium size (approx. 4” H x 6” W) printed circuit board that mounts inside the

enclosure. The field terminations are located along the lower edge of the board via pluggable

terminal block. All components and devices on the board are designed for extended temperature

and low drift operation.

2.3.7 Process Interface

An appropriate Process Interface is selected to suit the process/stack installation. The analyzer is

detachable from the process interface to facilitate Off-Line calibration and service.

Process Interface Options

There are several systems available as well as custom designs for specific applications.

ƒ

Flanged O-Ring Alignment comprises typical 2” or 3”

process flange with a large diameter O-Ring seal,

typically used for stack or inert applications that are

non-corrosive.

ƒ

Flanged Metal Bellows Seal comprises typical 2” or

3” process flange with a metal bellows seal and

external mechanical alignment system, typically used

when gas containment is important.

ƒ

Flow Cell may be used when the process gas has

been extracted or is used in a by-pass flow loop.

This allows for heat tracing (if necessary) and easy

introduction of both Zero and Span gases.

ƒ

Isolation Flanges are supplied with process windows

mounted in the flanges themselves typically for very

corrosive and/or high pressure applications.

ƒ

By-Pass Piping may be used when the process gas

line is of small diameter (typically <10”) and when no

suitable section of pipe work exists on this which the

analyzer can be mounted. By pass piping systems

may be provided or supplied at site. Heat tracing

may also be supplied or provided at site.

ƒ

Off Line Calibration Cell is used for off line

verification of the system. This cell is used to flow

Zero and Span gases when the analyzer is not

connected to the process.

Figure 8 - Process Interface Examples

Flanged with Metal Alignment

Typical Cross Pipe/Stack

Typical Flow Cell / By-Pass

Figure 13 - Process Interface Examples