Precautions Against Electrostatic Problems

The GC1000 system uses numerous IC components. When handling cards with IC

components mounted on them for maintenance or setting changes, take full precau-
tions against electrostatic problems.

These precautions are summarized below.

(a)

When storing or carrying cards, enclose them in a conductive bag or antistatic
bag. (Cards as shipped by Yokogawa are enclosed in a conductive bag or
antistatic bag labeled with cautions against electrostatic problems.)

(b)

Whenever mounting or demounting cards into or from a product, wear a wrist
strap grounded via a 1 M

resistance. Connect the wrist strap to any ground

terminal near the ground wire or to any unpainted part of the grounded frame.

Sheet

Gard

List strap

List strap

1M‰

F0003.EPS

Using a wrist strap and conductive sheet

Using a Conductive Sheet

(c)

When servicing cards on the bench, place them on a conductive sheet
grounded via a 1 M

resistance, wearing a wrist strap as in (2) above. Keep

easily-chargeable plastic materials away from the bench.

(d)

Never touch components mounted on the cards, the pattern side, connectors,
pin components, etc. with bare hands, unless using a wrist strap and a conduc-
tive sheet.

(e)

Wrist straps and conductive sheets are available from Yokogawa Engineering
Service (YSV).

3rd Edition : Aug. 23. 2006-00