4 communication input and output, Communication input and output -8, 1)analysis value data (component concentration) – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 65: 3)calibration factor, 4)error alarm

Advertising
background image

5-8

<Toc> <Ind>

<5. Actions of External Input and Output Signals >

IM 11B03A03-01E

3rd Edition : Aug. 23. 2006-00

5.4

Communication Input and Output

5.4.1

GC6 Type Output Data Format (Fixed to 45 Characters)

(1)Analysis value data (component concentration)

1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45

D S

,

,

%

PPM

A: CHL

A: CLL

A: RT

CR: Carriage return
LF: Line feed

Dummy

space

,

,

,

T

C

R

L

F

b

a

c

d

e

f

g

a: Stream number (01 to 24)

b: Component number (01 to 24)

c: Component concentration (without zero suppression)

d: Unit

e: Concentration error alarm (Upper limit = A:CHL, Lower limit = A:CLL)

f: Retention time (seconds)

g: Retention time error alarm

Note: The error alarm is output only if an error occurs. (Dummy space if no error occurs)

F0504.EPS

(2)Analysis value data (component ratio operation, linear polynormial operation)

1

2

3

4

5

6

7

8

9

10 11 12 13

15 16

17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43

44 45

D S

,

,

Dummy space

C

R

L

F

b

a

c

a: Stream number (01 to 24)

b: Component number (01 to 24)

c: Operation value (without zero suppression)

d: Unit

14

%

d

PPM

F0505.EPS

(3)Calibration factor

1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16

17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43

44 45

C C

,

,

,

COE

Dummy space

C

R

L

F

b

a

c

d

a: Standard sample (STD1=1, STD2=2)

b: Component number (01 to 24)

c: Calibration factor (without zero suppression)

d: Calibration factor error (only if an error occurs)

F0506.EPS

(4)Error alarm

• Memory pattern check error

• Watchdog timer

• Calibration out of range

• Detector 1 (or 2) calibration error

• Isothermal oven temperature error

• Power off

• Pressure switch 1 (or 2) OFF

• External contact inputs 1 to 8

• Communication error

• FID1 (or 2) extinguished

• Calibration repeatability error

1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16 17 18

19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43

44 45

A S

/

:

,

,

Dummy space

C

R

L

F

b

a

c

d

e

a: Month, b: Day, c: Hour, d: Minute

e: Error alarm types

MEM: Memory pattern check error

WDT: Watchdog timer

AD1: Detector 1 calibration error

AD2: Detector 2 calibration error

TMPH: Isothermal oven temperature error

CAR1: Pressure switch 1 OFF

CAR2: Pressure switch 2 OFF

EXT1: External contact

input 1

EXT2: External contact

input 2

EXT8: External contact

input 8

FLM1: FID1 extinguished

FLM2: FID2 extinguished

RPT: Calibration repeatability error

CAL: Calibration out of range

POF: Power off

NSD: Communication error (GC1000 - GCCU)

8 items in total

F0507.EPS

Note: For 3 characters, the 18th character is a dummy space.

Advertising