6 alarm processing, Alarm processing -22 – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 57

Advertising
background image

4-22

<Toc> <Ind>

<4. Outline of Software >

IM 11B03A03-01E

4.6

Alarm Processing

There are four alarm levels. For details of alarms, see the Alarm Message Manual, IM
11B03A03-06E.

Level 1

Alarms for system or hardware failures. When an alarm is generated, the alarm
condition is maintained until it is reset.

If a level-1 alarm is generated in the Run or Lab mode, the operation mode changes to
Stop at the end of the measurement being executed.

Level 2

Alarms for measuring conditions failure.

When an alarm is generated, the alarm condition is maintained until it is reset.

Component

Alarms when measurement results such as concentrations and retention times be-
come out of each specified range.

If an alarm is generated, the alarm condition is maintained until it is reset.

Level 3

Alarms for minor failures or information other than level-1 and level-2 alarms.

The alarm condition is not maintained.