Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 75

Advertising
background image

5-18

<Toc> <Ind>

<5. Actions of External Input and Output Signals >

IM 11B03A03-01E

(7)

Calibration factor

This register contains the calibration factor for each calibrated stream of the GC1000
Mark II. Since the factors are in the range of 0.000 to 9.999, each value is multiplied
by 1000 and displayed as integers, i.e., 0000 to 9999.

Table 5.5 Address Table (GC1000)

Description

Name

Address

Run command

Stop command

Time setting request

Calibration (validation)
command

Stream sequence select
command

Stream (Cont) select
command

001TT

0002P

00001

00002

Recieved the message by the master, the slave reset.

Same as above

Same as above

00003

0001P

Same as above P: Stream Sequence number (1 to 4)

Same as above P: 1 to 3 (Cal. 1 to 3), 4 to 6 (Val. 1 to 3)

TT: Stream number (01 to 31)

Analyzer normal

10001

Reset when the alarm status is read after alarm status change is read.

Analyzer error

10002

Analyzer status change

10003

Measuring

10004

Stop

10005

Maintenance

10006

Stream
sequence

1001P

Stream
(Cont)

10021

Calibration (Validation)

10022

Stream sequence

10023

Data update

101TT

Calibration factor update

102TT

Reset when the analysis value is read after data update is read. (TT:
Stream number)

Reset when the calibration factor is read after calibration factor
update is read. (TT: Stream number)

Data valid

11CCC

CCC: Component number

Alarm status

12AAA

AAA: Alarm number (001 to 249)

Year (40001), month/day (40002), hour (40003),
minute/second (40004)

Time setting request

40001 to 40004

Read only (Write disabled)
DDD(real format)5CCC*2-1(CCC: Component number)

Analysis value

41DDD

NN: Stream number +40, QQ: Peak number

Range select

4NNQQ

Stream number

30001

Starting peak number

301TT

Starting peak number of each stream (TT: Stream number)

Peak number of each stream (TT: Stream number)

Hour/Minute (TT: Stream number)

Seconds

Analysis value (CCC: component number)

Peak number

302TT

Sampling time

303TT

Retention time

32CCC

Calibration factor

33CCC

Analysis value

31CCC/-

31DDD

Analysis value (CCC: component number)
DDD(real format): CCC*2-1(CCC: Peak number)

P: Stream Sequence number (1 to 4)

T0505.EPS

3rd Edition : Aug. 23. 2006-00