<5. Actions of External Input and Output Signals >

5-13

IM 11B03A03-01E

3rd Edition : Aug. 23. 2006-00

Non-protocol output signal transmission timing

Send Data (SD)

Request to Send (RS)

Terminal Equipment Ready (ER)

Analysis data

AUTO mode analysis

completion point

+9V (RS232C)

Calibration factor

During AUTO mode calibration

+9V (RS232C)

Error alarm

When an alarm occurs (after transmission
completion if the component concentration
value or calibration factor is being transmitted)

+9V (RS232C)

(Always ON)

Processor power ON

+9V (RS232C)

(Transmission completion)

(Transmission completion)

F0516.EPS

I/O instruction format and communication control procedure for handshaking

C

R

L

F

# E

C

R

L

F

# F

C

R

L

F

# T

C

R

L

F

# B

#

#

W

C

R

L

F

# W

C

R

L

F

# A

C

R

L

F

# R

I/O instruction format

Communication control procedure

E: Data receiving request

F: Data transmission

completion

T: Input command prompt

B: Answerback

(command executed)

W: Answerback

(command not executed)

If the command is in
an incorrect format,
answerback is in the
format shown at right.)

A: Data transmission

request

R: Data retransmission

request

A

A

A

E

F

Data 1

Data n

Supervisory computer

GCCU

t (0.4 to 0.5 second)

t

t

t

t

t

t

Output upon analysis completion or occurrence of an error alarm
(after transmission completion if the component concentration value or calibration
factor is being transmitted)

● Normal communication

A

R

R

E

Data 1

Data 1

Data 1

Supervisory computer

GCCU

Retransmission

Note: If retransmission is requested, data can be transmitted any number of times. If the

total communication time exceeds one analysis cycle, the data is changed into
new data. (The limiter should be considered for the supervisory computer side.)

T

B or W

B or W

Command

1

5 seconds

Command

2

Supervisory computer

GCCU

● Input command

Note: If the sequence has stopped, input commands can be received continuously until

the start signal is received.

5 seconds

Up to 6 commands including those not executed

Output after
completion of
data transmission
or every 8 to 10
sec periodically

Re-prompted if there is no command.
If there is no command upon
re-prompting, it is assumed that there
is no command.

If there is no command
during this period, it is
assumed that there is no
command.

E

E

E

E

E

Supervisory computer

GCCU

● Error handling

3

seconds

3

seconds

E

3.2 seconds (for reference)

(No response from the computer)

The same
command is
repeated.

(No response from GCCU)

Output instruction

Input instruction

F0517.EPS

Note: If #ACRLF or #RCRLF is not received in 20 seconds or more after the data has been sent, the transmission of the

analysis data of that cycle will be aborted.