3) holding register (set data) – Yokogawa GC1000 Mark II Process Gas Chromatograph User Manual

Page 73

Advertising
background image

5-16

<Toc> <Ind>

<5. Actions of External Input and Output Signals >

IM 11B03A03-01E

(3) Holding Register (Set Data)

(1)

Time setting value

This is a set of four registers used by the DCS to set the device clock. When the time
setting request coil (address 00003 for GC1000 Mark II and GCIU, 00001 for the
analyzer server) is activated, these entries are used for the year, month/day, hour and
minute/second. Since all devices on the highway synchronize to the analyzer server
clock on a regular basis, the analyzer server clock should be regularly set.

F0518.EPS

Minute

Second

Hour

Month

Day

Year

Example : September 25, 1996, 15:23:10

07CC (hexadecimal)

0919 (hexadecimal)

000F (hexadecimal)

170A (hexadecimal)

Figure 5.4

Time Register Configuration

(2)

Range setting command

This address allows the range of the component indicated by the specified stream
number and peak number to be changed on the GC1000 Mark II.

(3)

Analysis value

These addresses display the same data as the analysis value in the input resister,
however values cannot be written to theses addresses. Only these addresses
support real number (floating point) form.

3rd Edition : Aug. 23. 2006-00