<4. Outline of Software >

4-1

IM 11B03A03-01E

3rd Edition : Aug. 23. 2006-00

4.

Outline of Software

4.1

Status and Operation Mode

The GC1000 Mark II has the following statuses:

Process: Normal measurement, calibration and validation

Manual:

Manual operation

Lab:

Measurement with automatic peak detection, like a laboratory GC

The GC1000 Mark II has the following operation modes:

Stop:

Mode in which the measurement stops

Run:

Mode in which the measurement is running

Pause:

Mode in which the measurement pauses

Figure 4.1 shows the transitions of statuses and operation modes.

Pause mode

Run mode

Run mode

Stop mode

Stop mode

Process

Manual

Run command

Run

command

Run

command

Stop command

Change the status

Change the status

Stop

command

Stop command

Pause command

Pause
command

Run mode

Stop mode

Lab

Run

command

Stop

command

F0401.EPS

Figure 4.1

Status and Operation Mode Transition Diagram

When the LCD panel on the GC1000 Mark II or the LCD emulator on a personal computer
(PC) connected with the GC1000 Mark II is used, by changing the operation panel to the
manual operation panel or lab operation panel, the status will change from Process to
Manual or Lab. Likewise, by changing the manual operation panel or lab operation panel
to the operation panel, the status will change from Manual or Lab to Process.

When the GCMT on a PC connected with the GC1000 Mark II is used, the status can be
changed through command button operation or menu-driven operation.

The following pages describe each status and operation in detail.