• The first temperature gradient/first control temperature/first temperature time (only for the
GC1000D/GC1000T/GC1000W)

• The second temperature gradient/second control temperature/second temperature
time (only for the GC1000D/GC1000T/GC1000W)

• The third temperature gradient/third control temperature /third temperature time
(only for the GC1000D/GC1000T/GC1000W)

• Initial control pressure

Displayed when EPC is installed that uses a pressure program. The pressure at the
start of analysis should be specified.

• Initial pressure time

Displayed when EPC is installed that uses a pressure program. The time for holding
the initial control pressure should be specified.

• First pressure gradient/first control pressure/first control time (only when EPC is
installed)

• Second pressure gradient/second control pressure/second control time (only when
EPC is installed)

• Third pressure gradient/third control pressure/third control time (only when EPC is
installed)

In the GC1000D/GC1000T/GC1000W, three stages of temperature program can be speci-
fied. Figure 4.3 shows the operations for temperature programs.

Initial

temperature

time

Analysis cycle

Analysis end

Analysis start

Time

Heater on

First

temperature

time

Second

temperature

time

Third

temperature

time

End of peak detection

time

Heater off

Third

temperature

gradient

Second

temperature

gradient

First

temperature

gradient

Temperature

Third control

temperature

Second control

temperature

First control

temperature

Initial control

temperature

F0403.EPS

Figure 4.3

Operation for Temperature Programs

3rd Edition : Aug. 23. 2006-00