3-4 setting of tags and addresses, 3-4 setting of tags and addresses -9 – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 14

Advertising
background image

IM 12A00A01-61E

Foundation Fieldbus 3-9

3-9

LI100

LIC100

FIC100

FC100

FI100

Commu-

nication

Schedule

OUT

IN

OUT

CAS_IN

BKCAL_OUT

BKCAL_IN

BKCAL_IN

BKCAL_OUT

IN

Unscheduled

Communication

Scheduled
Communication

Macrocycle (Control Period)

Figure 3.8 Function Block Schedule and Communication Schedule

When the macrocycle is set to more than 4 seconds, set the following intervals to be more than 1% of the
macrocycle.
- Interval between “end of block execution” and “start of sending CD from LAS”
- Interval between “end of block execution” and “start of the next block execution”

“end of block execution” and “start of the next block execution”

3-3-4 Setting of Tags and Addresses

Setting of Tags and Addresses

This section describes the steps in the procedure to set PD Tags and node addresses in the EXA.
There are three states of Fieldbus devices as shown in Figure 3.9, and if the state is other than SM_
OPERATIONAL state, no function block is executed. EXA must be transferred back to this state after a tag
or address is changed.

Tag clear

Tag setting

Address clear

Address setting

UNINITIALIZED

(No tag nor address is set)

INITIALIZED

(Only tag is set)

SM_OPERATIONAL

(Tag and address are retained, and

the function block can be executed.)

Figure 3.9

Status Transition by Setting PD Tag and Node Address

EXA has a PD Tag and node address that are set upon shipment from the factory unless otherwise speci-
fied. To change only the node address, clear the address once and then set a new node address. To set
the PD Tag, first clear the node address and clear the PD Tag, then set the PD Tag and node address
again.

Devices whose node address was cleared will await the default address (randomly chosen from a range
of 248 to 251, or from hexadecimal F8 to FB). It is necessary to confirm the device ID in order to correctly

order to correctly

specify the device. The device ID of the EXA is 594543083xxxxxxxxx. (The xxxxxxxxx at the end of the
above device ID is a total of 9 alphanumeric characters. The characters displyed on the screen when each
instrument is powered on, correspond to the xxxxxxxxx and if necessary, should be recorded.)

Advertising