4. function block parameters and methods, 4-1. physical block parameters, 4-2. analog input block parameters – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 66: 4. function block parameters and methods -8

Advertising
background image

4-8 Profibus

IM 12A00A01-61E

4-4. Function block parameters and Methods

4-4-1. Physical Block Parameters

Parameter

Default

Alternatives

R/W

Slot, Index

Data Type

(byte,bit)

(bytes)

Software revision

“R3.01”

R

0,24

visible string (16)

Hardware revision

“R3.01”

R

0,25

visible string (16)

Device manufacturer ID

37hex (Yokogawa)

R

0,26

unsigned16 (2)

Device ID

PH202 “5945430850”

R

0,27

visible string (16)

SC202 “5945430851”

R

0,27

visible string (16)

ISC202 “5945430852”

R

0,27

visible string (16)

DO202 “5945430853”

R

0,27

visible string (16)

Device serial number

“00000000U9313508” (example)

R

0,28

visible string (16)

Diagnosis

for details see .gsd file

R

0,29

bit string (4)

Diagnosis extension

for details see .gsd file

R

0,30

bit string (6)

Diagnosis mask

R

0,31

bit string (4)

Diagnosis mask extension

R

0,32

bit string (6)

Device certifiacation

R

0,33

visible string (32)

Write locking

Write enable (2457)

Write disable (0)

R/W

0,34

unsigned16 (2)

Factory reset

0

factory reset (1), warm start (2506),

reset bus address to 126 (2712)

R/W

0,35

unsigned16 (2)

Descriptor

YOKOGAWA PROFIBUS

-PA ANALYZER

R/W

0,36

visible string (32)

Device message

YOKOGAWA PROFIBUS

-PA ANALYZER

R/W

0,37

visible string (32)

Device install date

“”

R

0,38

visible string (16)

Ident number selector

manufacturer specific

ident no (1)

profile specific ident number (0)

R/W

0,40

unsigned8 (1)

Device configuration

“”

R

0,52

visible string (32)

Init state

run (2)

R/W

0,53

unsigned8 (1)

Device state

run (2)

R/W

0,54

unsigned8 (1)

Global status

0

bit0 - failure, bit1 - maintenance requested,

bit2 - function check, bit3 - limits exceeded,

bit4~15 - reserved

R

0,55

bit string (2)

4-4-2. Analog Input Block Parameters

Parameter

Default

Alternatives

R/W

Slot, Index

Data Type

(byte,bit)

(bytes)

AI1, AI2, AI3

Static Revision No.

R

1,17

unsigned16 (2)

Channel

PH202: pH(284)

Temp (298), ORP/rH(305)

SC202: SC1(284)

Temp (292), SC2(303), Conc(302)

ISC202: SC1(284)

Temp (292), SC2(303), Conc(302)

DO202: DO(284)

Temp (299), %Sat(306), sensor current(305) R/W

1,30

unsigned16 (2)

Linearization type

Non linearization (0)

R/W

1,29

unsigned 8 (1)

Unit

R

1,28

DS-36 (11)

Decimal point

1

R/W

1,28

DS-36 (11)

Filter Time Const

R/W

1,32

float (4)

Process Value Scale

Lower Value

R/W

1,27

float[2] (8)

Upper Value

R/W

1,27

float[2] (8)

Output Scale

Lower Value

R/W

1,28

DS-36 (11)

Upper Value

R/W

1,28

DS-36 (11)

Output Limits

Upper Limit Alarm

Maximum of float

R/W

1,37

float (4)

Upper Limit Warning

Maximum of float

R/W

1,39

float (4)

Lower Limit Warning

Minimum of float

R/W

1,41

float (4)

Lower Limit Alarm

Minimum of float

R/W

1,43

float (4)

Limit Hysteresis

0.5 %

R/W

1,35

float (4)