Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 38

Advertising
background image

IM 12A00A01-61E

Foundation Fieldbus 3-33

3-33

Index Parameter name

Factory

Valid

Description

Default

Range

2029 SECONDARY_VALUE_UNIT °C

°C, °F

Temperature unit

2030 SENSOR_TEMP_COMP

automatic

Off, manual,

Select off when no temperature compensation is required.

automatic

Select manual when no temperature element is available and the

temperature is stable and select auto when a temperature element

is available

pH: man + auto

orp: off + auto

2031 SENSOR_TEMP_MAN_VALUE 25

-30 to 140°C,

manual temperature value

-20 to 280°F

2032 SENSOR_TYPE_TEMP

Pt1000

Pt1000, Pt100,

Temperature element used:

5k1, 3kBalco,

8k55, 350,

NTC10k, 6k8

2033 SENSOR_CONNECTION

2

2

Only 2-wire connections supported

_TEMP

2034 TERTIARY_VALUE_TYPE

None

None, ORP, rH

When a metal electrode is used in combination with a glass and

reference electrode one has the possibility to measure a tertiary

value:

2035 TERTIARY_VALUE

-

-1500 to 1500 mV, The third value is expresed in mV or rH units

0 to 55 rH

2036 TERTIARY_VALUE_RANGE -1500

-1500 to 1500 mV, The high and low range limits for the tertiary value.

to 1500

0 to 55 rH

2037 TERTIARY_ZERO

0

-120 to 120mV

Redox offset can be set in mV.

2038 GLASS_IMPEDANCE

-

1MΩ to 2 GΩ (HI) Value that indicates the state of the glass membrane.

1kΩ to 1MΩ (LO)

Depending on the jumper setting the range is set to High or Low.

2039 REFERENCE_IMPEDANCE -

1MΩ to 2 GΩ (HI) Value that indicates the state of the reference system.

1kΩ to 1MΩ (LO)

Depending on the jumper setting the range is set to High or Low.

2040 ALARM_SUM

-

2041 DEV_ALARM

-

Device Alarm is used to give the status of the transmitter. Device_

status in RB displays the device Alarms.

2042 LOGBOOK1_RESET

-

Reset the pointer to the first (oldest) event in logbook 1.

2043 LOGBOOK1_EVENT

-

Event whereto the pointer is referenced. When parameter is read,

the pointer is increased by one.

2044 LOGBOOK2_RESET

-

Reset the pointer to the first (oldest) event in logbook 2.

2045 LOGBOOK2_EVENT

-

Event whereto the pointer is referenced. When parameter is read,

the pointer is increased by one.

2046 LOGBOOK_CONFIG

-

Per event one can decide whether it should be logged and in which

logbook (1 or 2) it should be logged.

2047 TEST_1

-

Service parameter

....

-

2048 to 2058 are, like 2047 and 2059, service parameters

2059 TEST_13

-

Service parameter

2060 STABLE_TIME

5

5.0 to 30.0

Stability criteria used during automatic calibration.

2061 STABLE_VALUE

0.02

0.01 to 1.0

Stability criteria used during automatic calibration.

2062 CALL_MAINT_TIME_

COUNTDOWN

250

1 to 250

The remaining days till maintenance is required.

2063 CALL_MAINT_TIME_RELOAD

250

1 to 250

The interval of the maintenance timer.

2064 INPUT_1_IMPEDANCE_LO_LIM

1.00E6

100.0 to 1.0E9

The low limit of the input 1 impedance.

2065 INPUT_1_IMPEDANCE_HI_LIM

1.00E9

100.0 to 1.0E9

The high limit of the input 1 impedance.

2066 INPUT_2_IMPEDANCE_LO_LIM

100

100.0 to 1.0E9

The low limit of the input 2 impedance.

2067 INPUT_2_IMPEDANCE_HI_LIM

200000

100.0 to 1.0E9

The high limit of the input 2 impedance.

2068 BUFFER1_ID

4

0 to 9

The ID of buffer one, used during automatic calibration.

2069 BUFFER1

4.0

-2.0 to 16.0

The first user defined calibration buffer.

2070 BUFFER2_ID

7

0 to 9

The ID of buffer two, used during automatic calibration.

2071 BUFFER2

7.0

-2.0 to 16.0

The second user defined calibration buffer.

2072 BUFFER3_ID

9

0 to 9

The ID of buffer three, used during automatic calibration.

2073 BUFFER3

9.0

-2.0 to 16.0

The third user defined calibration buffer.

2074 TEMPERATURE_COEFFICIENT

0.0

-1.0 to 1.0

The temperature coefficient of the primary value.

-100.0 to 100.0