6-3-2 transducer block sc202, 6-3-2 transducer block sc202 -35 – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 40

Advertising
background image

IM 12A00A01-61E

Foundation Fieldbus 3-35

3-35

3-6�-3-2 Transducer Block SC202

Index Parameter name

Factory

Valid

Description

Default

Range

2000 BLOCK HEADER

TAG: “TB”

General information about thefunction block

2001 ST_REV

-

-

The revision level of the static data associated with the function

block. The revision value will be incremented each time a static

parameter value in the block is changed

2002 TAG_DESC

“ “

The user description of the intended application of the block

2003 STRATEGY

1

The strategy field can be used to identify grouping of blocks.

This data is not checked or processed by the block

2004 ALERT_KEY

1

The identification number of the plant unit. This information may be

used in the host for sorting alarms, etc.

2005 MODE_BLK

AUTO

The actual, target, permitted, and normal modes of the block

2006 BLOCK_ERR

-

This parameter reflects the error status associated with a block. It is

a bit-string, so that multiple errors can be shown

2007 UPDATE_EVT

-

The alert is generated by any change to the static data.

2008 BLOCK_ALM

-

The block alarm is used for all configuration error, hardware

connection failure or system problems in the block. The cause of

the alert is entered in the subcode field. The first alert to become

active will set Active status in Status attribute.

2009 TRANSDUCER_DIRECTORY

A directory that specifies the number and

starting indices of the transducers.

2010 TRANSDUCER_TYPE

Standard

conductivity

Conductivity transducer block

Conductivity

2011 XD_ERROR

-

The error code in transducer:

No failure,

Electronics failure,

I/O failure,

Mechanical failure

2012 COLLECTION_DIRECTORY -

A directory that specifies the number, starting indices and DD item Ids

of the data collection in each transducer within a transducer block

2013 PRIMARY_VALUE_TYPE

Conductivity conductivity /

Type of measurement represented by primary value

resistivity

2014 PRIMARY_VALUE

-

0 to 2 S/cm

Primary value of the instrument is Conductivity

2015 PRIMARY_VALUE_RANGE

0 to 2 S/cm 0 to 2 S/cm

The range of the instrument (can not change)

2016 SENSOR_CONST

0.1

0.005 to 50 cm

-1

The conductivity cell has a specific cell constant determined by the

dimensions of the cell

2017 CAL_POINT_HI

1.999

0 to 2 S/cm

Highest calibration point

2018 CAL_POINT_LO

0

0 to 2 S/cm

Lower calibration point

2019 CAL_MIN_SPAN

0,0001

> 0,0001 S/cm

Minimum span between two calibration points

2020 SENSOR_CAL_METHOD

-

1point, 2point

not used

2021 SENSOR_CAL_DATE

-

till 2104

Date the sensor was last calibrated

2022 SECONDARY_VALUE

-

-20 to 250 ºC ,

Temperature value

0 to 500ºF

2023 SECONDARY_VALUE_UNIT ºC

ºC, ºF

Temperature unit

2024 SENSOR_TEMP_COMP

automatic

automatic

2025 SENSOR_TEMP_MAN_VALUE -

-

No manual temp. value can be set. Always Automatic

2026 SENSOR_TYPE_TEMP

Pt1000

Pt1000, Pt100,

Temperature element used:

Ni100, Pb36, 8k55

2027 SENSOR_CONNECTION_

2

2

Only 2-wire connections supported

TEMP

2028 SENSOR_TYPE_COND

contact

2-electrode,

Either 2-electrode or 4-electrode contacting

2-electrode 4-electrode

conductivity cell can be selected

2029 SENSOR_OHMS

-

Actual cell resistance

2030 XD_MAN_ID

“ “

2031 TEMPERATURE_COEFF

2.1

0 to 3.5%/ºC (%/ºF) Process temperature compensation factor

2032 CONCENTRATION

-

-

Conductivity combined with temperature can be directly related to

the concentration. Concentraion is expressed in percentage

2033 TERTIARY_VALUE

-

0 to 2 S/cm

Second compensated conductivity value