5-2 APPENDIX 1. LINK MASTER FUNCTIONS

IM 12A00A01-61E

A1-3 Transfer of LAS
There are two procedures for an LM to become the LAS:
• If the LM whose value of [V(ST)3V(TN)] is the smallest on a segment, with the exception of the current

LAS, judges that there is no LAS on the segment, in such a case as when the segment has started up or
when the current LAS has failed, the LM declares itself as the LAS, then becomes the LAS. (With this
procedure, an LM backs up the LAS as shown in the following figure.)

• The LM whose value of [V(ST)3V(TN)] is the smallest on a segment, with the exception of the current

LAS, requests the LAS on the same segment to transfer the right of being the LAS, then becomes the
LAS.

LM

Node address:

0x15

SlotTime = 5

LM

Node address:

0x16

SlotTime = 5

Node address: 0x14
SlotTime = 5

Basic device

Node address:

0xF1

Basic device

Node address:

0xF2

Basic device

Node address:

0xF3

Basic device

Node address:

0xF4

LAS

LAS

In the event that the current LAS in
this segment (node address 0x14)
fails, the LM with the address of 0x15
takes its place to become the LAS.

LM

Figure A1-2. Backup of LAS

To set up a EXA202 as a device that is capable of
backing up the LAS, follow the procedure below.

NOTE: When changing the settings in a EXA202,

add the EXA202 to the segment in which an
LAS is running. After making changes to
the settings, do not turn off the power to the
EXA202 for at least 60 seconds.

(1) Set the node address of the EXA202.

In general, use an address from 0x14 to
[V(FUN) – 1].

Not used

LM device

Not used

Basic device

Default address

Portable-device address

V (FUN)

V (FUN) + V (NUN)

V (NUN)

0xFF

0xFC

0xFB

0xF8

0xF7

0x00

0x0F

0x10

0x13

0x14

Bridge device

Figure A1-3. Node Address ranges

(2) In the LAS settings of the EXA202, set the

values of V(ST), V(MRD), and V(MID) to
the same as the respective lowest capability
values in all the devices within the segment.
An example is shown below.

Sub-

index

Element

Device

1

Device

2

Device

3

Description

1

3

6

SlotTime

MaxResponse
Delay

MinInterPdu
Delay

4

3

4

8

6

8

10

3

12

20

5

10

Capability value
for V(ST)

Capability value
for V(MRD)

Capability value
for V(MID)

DlmeBasicInfo (EXA202 Index 361 (SM))

EXA202

In this case, set SlotTime, MaxResponseTime,
and MinInterPduDelay as follows:

Subindex

Element

ConfiguredLinkSettingsRecord
(EXA202 Index 369 (SM))

Description

Setting

(Default)

V (ST)

V (MRD)

V (MID)

20

6

12

(4095)

(5)

(12)

SlotTime

MaxResponseDelay

MinInterPduDelay

1

3

6