Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 44

Advertising
background image

IM 12A00A01-61E

Foundation Fieldbus 3-39

3-39

Index Parameter name

Factory

Valid

Description

Default

Range

2035 COMP_METHOD

NaCl

NaCl, TC, matrix

Method of process temperature compensation for the primary value

2036 COMP_MATRIX_SEL

H

2

SO

4

H2SO4, 0 -100ºC, 0 - 5%, When matrix compensation is required one can

H2SO4, 0 -100ºC, 2.5 - 25%, make a selection out of 8 predefined matrices and
HCl, 0 - 60ºC, 0.5 - 5%,

one user definable matrix

HNO3, 0 - 80ºC, 0.5 - 5%
HNO3, 0 -80ºC, 2.5 - 25%
NaOH, 0 -100ºC, 0.5 - 5%

NaOH, 0 -100ºC, 0.5 - 15%

User programmable matrix

2037 TERTIARY_COMP_METHOD NaCl

NaCl, TC, matrix

Method of process temperature compensation

for the second conductivity value

2038 TERT_TEMPERATURE

2.1

0 ~ 3.5 %/ºC

Process temperature compensation factor for the

_COEFF

second conductivity value

2039 ALARM_SUM

-

2040 DEV_ALARM

-

Device Alarm is used to give the status of the analyser.

See separate table for error messages

2041 LOGBOOK1_RESET

Idle

Idle, Reset

Reset the pointer to the first (oldest) event in logbook 1

2042 LOGBOOK1_EVENT

-

Event whereto the pointer is referenced. When parameter

is read, the pointer is increased by one.

2043 LOGBOOK2_RESET

Idle

Idle, Reset

Reset the pointer to the first (oldest) event in logbook 2

2044 LOGBOOK2_EVENT

-

Event whereto the pointer is referenced. When parameter

is read, the pointer is increased by one

2045 LOGBOOK_CONFIG

-

Per event one can decide whether it should be logged and

in which logbook (1 or 2) it should be logged

2046 TEST_1

-

....

-

2047 to 2057 are, like 2046 and 2058, service parameters

2058 TEST_13

-

2059 CALIB_SENSOR_CONST

1.88

0.2~19.99 cm

-1

The calibrated cell constant [1/cm].

2060 MATRIX_TEMP_RANGE

-20~140ºC, 0~280ºF

The temperature values of the user defined matrix.

2061 SOLUTION_1

0~1.999 S/cm

The conductivity values of solution 1.

2062 SOLUTION_2

0~1.999 S/cm

The conductivity values of solution 2.

2063 SOLUTION_3

0~1.999 S/cm

The conductivity values of solution 3.

2064 SOLUTION_4

0~1.999 S/cm

The conductivity values of solution 4.

2065 SOLUTION_5

0~1.999 S/cm

The conductivity values of solution 5.

2066 CONCENTRATION_

disabled

0=Not initialized, 1=disabled,

Enabling/Disabling of concentration measurement.

MEASUREMENT

2=enabled

2067 CONCENTRATION_0

0

0~100

The concentration [%] corresponding to the 0 percent table value

2068 CONCENTRATION_100

100

0~100

The concentration [%] corresponding to the 100 percent table value

2069 CONC._TABLE_LOW

0~1.999 S/cm

The first part of the concentration table (0 till 30 percent).

2070 CONC._TABLE_MID

0~1.999 S/cm

The second part of the concentration table (35 till 65 percent).

2071 CONC._TABLE_HIGH

0~1.999 S/cm

The third part of the concentration table (70 till 100 percent).

2072 E5_LIMIT

3 (3 S)

0~5 S

The upper range accuracy limit of the analyzer [Siemens].

2073 E6_LIMIT

5e-6 (5

μS) 0~5 S

The lower range accuracy limit of the analyzer [Siemens].

2074 DISPLAY_RESOLUTION

1

0 to 8

Display resolution, 0, Not initialized (can not change)
1, Auto Ranging, 2, x.xxx μS/cm, 3, xx.xx μS/cm,
4, xxx.x μS/cm, 5, x.xxx mS/cm, 6, xx.xx mS/cm,
7, xxx.x mS/cm, 8, xxxx mS/cm

2075 PASSCODE_MAINT

0

0, 111, 333, 777,

Passcode used to protect the maintenance menu

888, 123, 957, 331, 546, 847

2076 PASSCODE_COMM

0

0, 111, 333, 777,

Passcode used to protect the commissioning menu.

888, 123, 957, 331, 546, 847

2077 PASSCODE_SERVICE

0

0, 111, 333, 777,

Passcode used to protect the service menu.

888, 123, 957, 331, 546, 847

2078 ERROR_CONFIG*

0=soft, 1=hard

Soft/Hard fail configuration (checked = hard unchecked = soft).

2079 CONFIGURATION*

autoreturn on

Instrument specific configuration. See below for correct

bitnr. setting

2080 TRANSMITTER_TIME

Transmitter time.”

2081 TEST_14

Service parameter.